Exposure, Density and the Logarithm
Two quantities have to be measured before anything else in sensitometry can be said: how much light went in, and how dark the film came out. Neither is difficult. What makes them usable together is a piece of arithmetic you already have — the base-ten logarithm — applied to light instead of to hydrogen ions.
By the end of this page you should be able to take a lamp, a timer and a step wedge, and write down the exposure every step of that wedge delivered; take a negative and a light meter, and write down a density; and read the axes of a manufacturer’s published curve without guessing what the numbers mean.
Exposure is a product, not a setting
Section titled “Exposure is a product, not a setting”In ordinary speech “exposure” means a shutter setting, or a frame of film, or the act of taking a picture. In sensitometry it is none of those. Exposure, written H, is illuminance multiplied by time.
E is the illuminance falling on the emulsion — how brightly lit the film’s surface is — measured in lux. t is the time that illuminance lasts, in seconds. Their product H is therefore in lux-seconds, and that is the unit along the bottom of every characteristic curve in this course. The photographic convention is fixed by ISO 6, cited here by number and nowhere quoted; Kodak’s sensitometry workbook states the same relation as Exposure = Illuminance × Time and works its examples in millilux and millilux-seconds, a thousandth of a lux, because camera-film exposures in lux alone are inconveniently small numbers.
Kodak’s own worked example is a good one to copy. An illuminance meter reads 75 millilux at the film plane and the shutter is open for one fifteenth of a second. Then H = 75 ÷ 15 = 5 millilux-seconds. Nothing more to it than that.
What a lux actually is, since the unit does some quiet work here. It is one lumen per square metre: luminous flux arriving per unit area, weighted by the sensitivity of the human eye across the spectrum. Two things follow. First, lux is a photometric unit, not a radiometric one — it counts light the way an eye counts it, not the way a photon detector does, so a source rich in blue and poor in green delivers more photographic effect per lux to a blue-sensitive emulsion than the number suggests. That is exactly why the standards specify an illuminant as well as a quantity, a point the speed lesson returns to. Second, it is per square metre, so it does not depend on how big the illuminated patch is; a lux meter held at the baseboard reads the same whether the enlarger is projecting a 35 mm frame or a whole sheet, provided the light falling on the cell is the same.
The reason the product is what matters — rather than the brightness or the duration separately — is the reciprocity law: half the light for twice the time is assumed to produce the same result. That assumption is what allows a single number on the horizontal axis to stand for an unlimited number of different lamp-and-timer combinations. It is also only approximately true, and it fails at both extremes. Part IV owns the mechanism and Part VI owns the correction; this page simply notes which assumption it is standing on.
The logarithm, extended rather than re-taught
Section titled “The logarithm, extended rather than re-taught”Part III taught the base-ten logarithm once, for pH, and said explicitly that Part XIII would build on it rather than repeat it. So, in one sentence: the base-ten logarithm of a number is the power of ten that makes it, log 1000 = 3 and log 0.001 = −3, and going back the other way is the antilog, 10 raised to that power.
What pH did not need, and sensitometry does, is the behaviour of logarithms under multiplication.
Read that as: multiplying quantities means adding their logarithms. Every convenience in the rest of this part is a consequence. Exposure is a product, E × t, so on a log axis it is a sum: doubling the time moves you a fixed distance to the right no matter what the illuminance was. Density is a logarithm of a ratio, so stacking two filters adds their densities. And a factor of two — one stop — is always the same distance, because log 2 = 0.301, whatever it is a factor of two of.
0.30 is the number to memorise. One stop, one step, one doubling: 0.30 in log exposure. Three of those is 0.90, close enough to 1.0 that a factor of ten and three stops are nearly the same move, which is why log scales and stop scales get confused. They are not the same: eight is not ten. Ten stops is 3.01 in log H; ten times the light is 1.00.
Photographers count in factors of two, so their natural base is 2 and their natural unit is the stop. Sensitometry counts in factors of ten, because that is what makes density and exposure share one scale. Converting is a single multiplication:
If you want the stop count directly from a ratio rather than from a log difference, the base-two logarithm does it, and it is one division away from the base-ten one you already have:
So a subject spanning 128 to 1 is log 128 = 2.107 in log terms and 2.107 ÷ 0.301 = 7.0 stops. There is no separate arithmetic to learn; there is one logarithm and a conversion factor.
The antilog is the return trip, and you will need it more often than you expect, because a speed number is an antilog and a transmittance is an antilog. It is the 10^x key on any calculator: if log H = −2.10 then H = 10^(−2.10) = 0.0079 lux-seconds. A useful sanity check while you are learning is that the whole-number part of a logarithm tells you the decimal place and the fractional part tells you the digits: 10^2.30 = 200 and 10^(−0.70) = 0.20 share the same fractional part, 0.30, and therefore the same leading digit 2.
Kodak’s workbook, incidentally, prefers step to stop for a factor of two in exposure, reserving “stop” for a change made at the lens aperture. The distinction is worth knowing when you read older literature; this course uses “stop” for the factor of two and says so here once.
Why both axes are logarithms
Section titled “Why both axes are logarithms”There are two good reasons and one that gets over-claimed.
Compaction. A subject on a bright day spans a luminance range of thousands to one, and a step wedge spans a thousand to one in transmission. An arithmetic axis from 1 to 1000 wastes nine tenths of its length on the top decade and crushes everything interesting into the last millimetre. A logarithmic axis from 0 to 3 gives each decade the same space. Kodak’s workbook makes exactly this argument and no other structural one.
How large a range is that in practice? The workbook’s own latitude example takes a scene whose brightest important object is 60 times brighter than its darkest, which is log 60 = 1.8 in log exposure, and sets it against a film curve covering 3.0. Two numbers, one subtraction, and you know there is 1.2 in log H of room to place the scene — four stops, split between the under- and overexposure ends. That calculation is the whole of exposure latitude, and it is only that easy because both quantities are logarithms. This course does not quote a figure for a “typical” scene range, because it has not verified one against a survey it can cite; 60 to 1 is Kodak’s worked example, not a measured average, and it is used here as an example and nothing more.
Additivity. This is the reason that actually changes what you can do. Because densities add, a ladder of equal density steps is a ladder of equal exposure intervals, and you can compute the exposure at every step of a wedge by subtraction instead of by dividing repeatedly. That property is what makes a step wedge an instrument rather than a set of grey patches.
And the one to be careful with. It is often said, and Kodak’s workbook says it in scare quotes, that the eye works “logarithmically” — that 0, 1 and 2 density units look like equal steps of brightness where 0.1, 0.2 and 0.3 in transmittance would not. Stated as an approximation over a limited range that is a reasonable working claim, and it does describe why a print’s tonal scale looks even. Stated as a law of vision it is more than the evidence in this course’s corpus supports, and nothing on these pages depends on it. The axes are logarithmic because of compaction and additivity, which are arithmetic facts; the eye is a bonus, not a premise.
Transmittance, opacity, density
Section titled “Transmittance, opacity, density”Three names for one measurement, which is why the definitions get muddled.
Transmittance, T, is the fraction of the light that gets through: shine 100 lux at a negative, measure 25 lux on the far side, and T = 0.25. It runs from 1 (nothing stopped) down towards 0.
Opacity, O, is its reciprocal, 1/T, so the same negative has an opacity of 4. Opacity answers “how many times more light must I put in to get one unit out?” — which is the question a printer is actually asking.
Density, D, is the base-ten logarithm of the opacity.
Worked at the values worth knowing by heart:
| Transmittance T | Light passed | Opacity O = 1/T | Density D = log₁₀ O |
|---|---|---|---|
| 1.0 | all of it | 1 | 0.00 |
| 0.5 | half | 2 | 0.30 |
| 0.25 | a quarter | 4 | 0.60 |
| 0.1 | a tenth | 10 | 1.00 |
| 0.01 | a hundredth | 100 | 2.00 |
| 0.001 | a thousandth | 1000 | 3.00 |
Two things fall out of that table. A density of 0.30 is one stop, exactly as a log exposure interval of 0.30 is one stop — the two axes are in the same currency, which is why a sensitometric plot is drawn with 0.30 of density occupying the same distance on the paper as 0.30 of log exposure. Kodak’s workbook gives that as a construction rule, and it is not cosmetic: the contrast-index construction on the third lesson of this part measures a distance along a sloping line and is simply wrong if the scales differ.
And the range of a real negative is smaller than beginners expect. Base-plus-fog sits near 0.10 to 0.20; a well-exposed pictorial negative’s highlights rarely pass 1.4 or so. The whole picture lives in about 1.2 density units — a factor of sixteen in transmitted light.
Transmittance against density
- T = 10 to the power minus D
Show the numbers behind this plot
| Series | Density D | Transmittance T (fraction of light passed) |
|---|---|---|
| T = 10 to the power minus D | 0.00 | 1.00 |
| T = 10 to the power minus D | 0.10 | 0.79 |
| T = 10 to the power minus D | 0.20 | 0.63 |
| T = 10 to the power minus D | 0.30 | 0.50 |
| T = 10 to the power minus D | 0.40 | 0.40 |
| T = 10 to the power minus D | 0.50 | 0.32 |
| T = 10 to the power minus D | 0.60 | 0.25 |
| T = 10 to the power minus D | 0.80 | 0.16 |
| T = 10 to the power minus D | 1.00 | 0.10 |
| T = 10 to the power minus D | 1.20 | 0.06 |
| T = 10 to the power minus D | 1.40 | 0.04 |
| T = 10 to the power minus D | 1.60 | 0.03 |
| T = 10 to the power minus D | 1.80 | 0.02 |
| T = 10 to the power minus D | 2.00 | 0.01 |
| T = 10 to the power minus D | 2.40 | 0.00 |
| T = 10 to the power minus D | 3.00 | 0.00 |
Densities add, and that is what a step wedge is
Section titled “Densities add, and that is what a step wedge is”Put two filters in the beam. The first passes a fraction T₁ of the light, the second passes T₂ of what reaches it, so the pair passes T₁ × T₂. Take logarithms and the multiplication becomes addition:
A step wedge is that property manufactured. A Stouffer T2115 carries 21 patches whose densities rise in nominal increments of 0.15 from about 0.05 to 3.05 — a total range of 3.00, which is a thousand to one in transmission, in 21 equal ratios. Each 0.15 step is half a stop. Contact-print it onto film under a uniform lamp and each patch delivers a known exposure, computed by subtraction rather than measured:
That single subtraction is the reason sensitometry can be done at home at all. You need to know the exposure at one place — under the clear edge of the wedge — and the wedge itself supplies every other value on the axis.
It also explains a piece of vocabulary you will meet in Part XIV. A wedge exposed all at once under one lamp is an intensity-scale exposure: every patch is exposed for the same length of time and differs only in how much light arrived. The alternative is a time-scale exposure, where one uniform patch of film is given a series of different durations, as in a printing test strip. The two are equivalent only where the reciprocity law holds, and where it does not they give measurably different curves — which is why a sensitometer is built around a wedge rather than a shutter, and why the lab in this part exposes for a few seconds rather than a few minutes.
Where each number on the exposure axis comes from
- Lamp and timer — measured once, as illuminance times time; here log H 3.35 in millilux-seconds
- Step wedge, 21 patches — nominal 0.15 apart, 0.05 to 3.05
- Film, emulsion up, in contact — each patch receives log H of the source minus that patch density
What density is physically
Section titled “What density is physically”Density is an optical measurement, but it is measuring something material, and the chain from one to the other is worth having straight.
Developed silver stops light in two ways: it absorbs it, and it scatters it. Absorption is the electrons in a metal particle taking energy out of the wave. Scattering is the same particle sending light off in some other direction, still in the world but no longer travelling towards your detector, which is lost light as far as the reading is concerned. That distinction looks like hair-splitting and is not: absorbed light is gone whatever instrument you use, while scattered light can be recovered by an instrument that collects widely enough — and that is the whole of the geometry problem two paragraphs below.
Hurter and Driffield set the whole thing out in 1890 in terms that are still recognisable — they used T for transparency and O for opacity, noted that T × O = 1, and defined density as the number of absorbing particles per unit area multiplied by an absorption coefficient. Two differences from modern practice are instructive.
They worked in natural logarithms. Their relation was D = loge O, not log₁₀ O. The modern base-ten convention makes a density of 1.0 mean exactly a tenth, which is far more useful, but it means an H and D density figure read from an 1890 paper is 2.303 times the modern one for the same negative. If you ever compare their numbers with a datasheet, that factor is why they look wrong.
They asserted that density is directly proportional to the silver per unit area and used it that way. For their materials and their purposes it served. The modern statement adds a term:
Covering power is the density obtained per unit mass of silver per unit area, and it is not a constant. Part IV derived it in the Nutting form and showed it goes as one over the particle size: the same mass of silver divided into smaller particles gives more density, which is why a filamentary developed grain is so efficient and why a solvent developer, which grows more compact particles, returns slightly less density for the same silver. So density is a measure of silver and of how that silver is laid down, and the course keeps density and silver mass as separate ideas for exactly that reason.
There is a third thing the number depends on, and it is not a property of the negative at all: the geometry of the measurement. Because a silver image scatters, the density you read depends on how much of the scattered light your instrument collects. Diffuse density collects it from all directions; specular or projection density collects only what carried straight on, and reads higher. ISO 5-2 — cited here by number and nowhere quoted — separates the two and specifies a geometry for each, which is why Kodak prints “Densitometry: diffuse visual” beside its published curves rather than leaving it to be assumed. The practical consequence is the Callier effect, and it is the reason one negative prints harder under a condenser enlarger than under a diffuser.
Stops, EV and log H
Section titled “Stops, EV and log H”Three scales, one quantity, and a conversion you will do constantly.
| Change | In stops or EV | In log H | As a factor |
|---|---|---|---|
| One third of a stop | 1/3 | 0.10 | 1.26 |
| Half a stop | 1/2 | 0.15 | 1.41 |
| One stop | 1 | 0.30 | 2 |
| Two stops | 2 | 0.60 | 4 |
| Three and a third stops | 3 1/3 | 1.00 | 10 |
| Ten stops | 10 | 3.01 | 1024 |
Exposure value, EV, is a scale in which one unit is one stop, so an EV difference converts to log H by the same 0.30. That matters when a spot meter is used as an interim densitometer, and it matters when a subject’s luminance range is measured: a scene reading EV 8 in the shadows and EV 15 in the highlights spans 7 stops, which is 7 × 0.30 = 2.1 in log luminance. Hold that number; the third lesson multiplies it by a slope to predict what the negative will do.
Be careful with EV in one respect. A difference in EV is unambiguous — it is a number of stops. An absolute EV number is tied to a metering convention and to a film speed setting, so two meters can report different absolute values for the same wall while agreeing perfectly about the seven stops between wall and window. Sensitometry only ever needs the difference, which is one less thing to calibrate.
Going the other way, density to per cent transmission, is the antilog:
| Density | Per cent transmission | Read as |
|---|---|---|
| 0.10 | 79 % | base-plus-fog on a clean film |
| 0.30 | 50 % | one stop |
| 0.45 | 35 % | a thin shadow |
| 0.75 | 18 % | a mid-tone on many negatives |
| 1.20 | 6.3 % | a well-placed highlight |
| 2.00 | 1.0 % | approaching blocked |
| 3.05 | 0.089 % | the last step of a 21-step wedge |
Running that table backwards is the skill that matters at the bench. A patch that visibly passes about a fifth of the light has a density near 0.7; a patch you can just see a bright lamp through has a density somewhere above 2. Neither is a measurement, but both are enough to tell you whether your exposure series has landed in the right part of the wedge before you commit a whole roll to it.
Reading a manufacturer’s axes
Section titled “Reading a manufacturer’s axes”Published curves use three different horizontal scales and rarely explain which.
Absolute log H in lux-seconds. Kodak prints its Tri-X and T-Max characteristic curves this way, with the axis running from about −4.0 up to 1.0, because a camera film needs only small fractions of a lux-second. Negative numbers on the axis are normal and mean nothing sinister: log H = −2.0 is 0.01 lux-seconds.
Millilux-seconds. Kodak’s teaching workbook uses these instead, which shifts every number up by exactly 3.00 and makes the axis positive. A speed point at log H −2.10 in lux-seconds is the same point as log H 0.90 in millilux-seconds. When you compare a worked example from the workbook with a curve from a datasheet, check the unit before you conclude anything.
Relative log exposure. ILFORD’s contrast-control literature plots its paper curves this way, with no absolute scale at all. That is entirely legitimate — the shape and the horizontal separation of the curves is the whole message — but it means no speed can be read from such a plot, only a comparison.
Alongside the axis, a serious published curve states what was done: Kodak’s carry the developer, the temperature, the agitation, the four development times and the densitometry condition, all on the plot. A curve without those is not a measurement, it is a picture of one.
Do the maths
Section titled “Do the maths”- Exposure H is illuminance times time, in lux-seconds. The product is what the emulsion responds to, and treating it as a single number assumes the reciprocity law, which is an approximation.
- Density is the base-ten logarithm of the reciprocal of transmittance. D 0.30 passes half the light, D 1.00 a tenth, D 3.00 a thousandth.
- Logarithms turn multiplication into addition, which is why densities add, why a step wedge is a ladder of equal exposure intervals, and why one stop is 0.30 wherever it appears.
- Both axes are drawn to the same scale, 0.30 of density occupying the same distance as 0.30 of log exposure, because the constructions in the next two lessons measure distances along a slope.
- Density measures silver and how the silver is divided, through covering power, and what the instrument reads also depends on its geometry.
- Check the unit on any published curve: lux-seconds, millilux-seconds and relative log exposure are all in use, and they differ by 3.00, by an arbitrary offset, or by everything.
Check your understanding
Sources for this page
8 cited · checked 2026-09-05
- 01Basic Photographic Sensitometry Workbook, publication H-740Eastman Kodak Company§ Names of Units - Exposure equals Illuminance times Time, with the worked example of 75 millilux for one fifteenth of a second giving 5 millilux-seconds; Density - transmission, opacity as the reciprocal of transmission, and density as the base-ten logarithm of opacity; Why Logs - compaction of the scale and the claim that the eye works logarithmically; Step Tablets - the 11-step and 21-step tablets, both about 0.05 to 3.05, at increments of 0.30 and 0.15; Figuring Exposure - the subtraction of filter and step-tablet densities from the log exposure of the source; Constructing the Curve - the rule that 0.30 in density must match 0.30 in log exposure on the paper; Glossary - the preference for step over stop as the general term for a factor of two in exposurekodak.com/content/products-brochures/Film/Basic-Photographic-Sensitometry-Workbook.pdftier 1, primary2026-09-05
- 02Memorial Volume containing an account of The Photographic Researches of Ferdinand Hurter and Vero C. Driffield, being a Reprint of their Published Papers, together with a History of their Early Work and a Bibliography of Later Work on the same subjectEdited by W. B. Ferguson, K.C., M.A., F.I.C., Hon. F.R.P.S., 1920§ Photochemical Investigations, the section defining transparency, opacity and density, in which T times O equals one, density is the number of absorbing particles per unit area multiplied by the coefficient of absorption, the three quantities are related by T equals e to the minus D, and the density is stated to be directly proportional to the amount of silver deposited per unit areaarchive.org/details/memorialvolumeco00hurtialatier 1, primary2026-09-05
- 03ISO 6:1993, Photography - Black-and-white pictorial still camera negative film/process systems - Determination of ISO speed, second edition, 1993-02-01ISO/TC 42, Photography, 1993§ Cited by number only, as the standard that fixes the photographic convention of exposure H in lux-seconds and its logarithm as the horizontal axis of a sensitometric curve; no part of it is quotediso.org/standard/3586.htmltier 1, primary2026-09-05
- 04ISO 5-2:2009, Photography and graphic technology - Density measurements - Part 2: Geometric conditions for transmittance density, fifth edition, 2009-12-01ISO/TC 42 Photography and ISO/TC 130 Graphic technology, joint working group, 2009§ Cited by number only, as the standard that separates diffuse from projection transmittance density and specifies a geometry for each; consulted in the publisher's free preview for title, edition and contents, and quoted nowhereiso.org/standard/52914.htmltier 1, primary2026-09-05
- 05Transmission Step WedgesStouffer Industries, doing business as Stouffer Graphic Arts§ Product table - the T2115 with 21 steps at a 0.15 density increment, described as half a stop, to a maximum density of 3.05; and the statement that the T2120CC and T1530CC are the parts calibrated against NIST Standard Reference Material 38120Cstouffer.net/TransPage.htmtier 1, primary2026-09-05
- 06KODAK PROFESSIONAL TRI-X 320 and 400 Films, publication F-4017Kodak Alaris Inc., 2016§ Characteristic Curves - the axes as printed, log exposure in lux-seconds running negative to about 1.0, density to 4.0, with the densitometry stated as diffuse visual and the process, temperature, agitation and four development times named on the plotbusiness.kodakmoments.com/sites/default/files/files/resources/f4017_TriX.pdftier 1, primary2026-09-05
- 07KODAK PROFESSIONAL T-MAX 100 Film, publication F-4016Kodak Alaris Inc., 2016§ Characteristic Curves - the same axes as the Tri-X sheet, log exposure in lux-seconds against density, with the densitometry stated as diffuse visualkodakprofessional.com/sites/default/files/wysiwyg/pro/resources/f4016_TMax_100.pdftier 1, primary2026-09-05
- 08Contrast Control for ILFORD MULTIGRADE Variable Contrast Papers, technical informationHARMAN technology Limited (ILFORD Photo), 2010§ The two emulsion diagrams, whose horizontal axis is labelled relative log exposure with no absolute scaleilfordphoto.com/wp/wp-content/uploads/2017/03/Contrast-control-for-Ilford-Multigrade.pdftier 1, primary2026-09-05
Formulas, hazard statements, historical dates and process descriptions on this page were checked against the sources above on the date shown. Safety data changes: obtain the current safety data sheet for the product you actually buy before you open it.