Experiment: Agitation, Stand and Semi-Stand Development
Everyone who develops film has an opinion about agitation, and almost nobody has a number. This session produces four numbers per regime and one more that nothing earlier in the course has measured at all: how even the development was, read off the film at twenty stated positions rather than judged by holding the negative up to a window. It also does something the enthusiasm around stand development rarely does — it says, before any film is spent, exactly what the published sources contain and what they do not.
Purpose
Section titled “Purpose”To develop matched strips under three agitation regimes at one dilution, time and temperature; to read contrast index, the relative speed point, maximum density and a compensation ratio off each; to measure each strip’s evenness as a density profile along its length; and to repeat the comparison on continuous-tone roll film, where the failure modes live.
Hypothesis. That reducing agitation lowers the upper-scale gradient more than the lower-scale one — so the compensation ratio defined below falls — while leaving the relative speed point inside the resolution of the apparatus, and that it raises the standard deviation of the density profile above the control’s floor by a factor the instrument can report. It fails if the compensation ratio does not move by more than the resolution sentence allows; if the speed point moves further than the gradient change accounts for; or if the least-agitated cell is as even as the control, which would mean either that this vessel is too small for the effect or that the effect is smaller than the practice claims.
Second hypothesis, the floor arm. That the same regime at a dilution putting the developing agent per unit film area below the manufacturer’s stated minimum gives a curve that looks compensating and is in fact starved. It fails if that cell’s upper scale is not depressed relative to the same regime run above the floor.
The control. The manufacturer’s specified agitation scheme at the manufacturer’s published time: ILFORD’s intermittent scheme, four inversions in the first 10 seconds and four again in the first 10 seconds of each further minute, at 16 minutes, 20 °C, Kodak D-76 diluted 1+3 — ILFORD’s published time for FP4 Plus at EI 125 in that developer at that dilution. It is run in both vessels this session uses, which is what ties the two arms together.
The one variable. The agitation regime: same emulsion batch, lamp, wedge, developer bottle, dilution, volume per unit area, vessel, temperature, time, fixer, wash and reader. Where a second variable has to move — in the floor arm and in the long-time cell — it is named in the cell’s own name, and the analysis never attributes those cells to agitation alone.
Learning objectives
Section titled “Learning objectives”By the end of the session and the reading that follows it you will be able to:
- state the transport mechanism agitation controls, and derive from it why the same local bromide accumulation is an adjacency effect at one scale and bromide drag at another;
- define compensation as a ratio of two average gradients before you have any data, and compute the smallest difference in it your densitometer can report;
- design a stand arm against the developing-agent floor rather than against a dilution ratio, and convert any volume, dilution and film area into millilitres of stock per 80 square inches;
- measure evenness as a density profile, with the control cell’s own profile as the floor, instead of describing it;
- say which claims about stand development your apparatus can settle, which it cannot, and which have no Tier 1 or Tier 2 evidence behind them at all.
Prerequisites
Section titled “Prerequisites”Designing an experiment that yields a number owns the method, and nothing here re-derives it. The development time and temperature series must come first, because its contrast-index-against-time curve turns a contrast difference into an equivalent time — the only fair currency for regimes that develop at different rates — and four curve families, measured carries the uncertainty budget this page inherits. The chemistry is in development kinetics, acutance, adjacency effects and compensating development and diffusion and osmosis; the two constructions in gamma, contrast index and average gradient and film speed and exposure index.
The hard prerequisites are the two certificates: the sensitometer’s and the densitometer’s. Without a repeatability figure for the densitometer there is no evenness measurement at all, because the quantity is a spread of readings and you could not say how much of it was the instrument.
Safety classification
Section titled “Safety classification”Level A. Standard home-darkroom controls, on a session that opens no jar of powder.
Every solution is diluted from something already made: the D-76 stock from Part VIII’s mixing lab, the fixer and wetting agent from bought concentrates. Under the course’s rubric that means gloves, eye protection whenever a solution is poured or agitated, dedicated labelled vessels and the ordinary ventilation of a wet bench — the controls Part IX’s test-negative lab set for the same operation.
One thing here is not ordinary. Two cells stand unattended for an hour in open cylinders, in a room dark for part of it, which is how alkaline solution gets knocked over or topped up by somebody else. Label both cylinders on the vessel before filling, stand them in a tray that holds their whole contents, and put them where nothing else has to reach across them.
What is not a hazard here, and why. No powder is weighed, so the inhalation route that makes metol and hydroquinone a Level B proposition in the mixing lab is not open in this session: both are present, but in solution at a working dilution, so the control is contact rather than airborne. Nothing is heated. No concentrated alkali is handled, because D-76’s alkali is borax already in solution. A fixed and washed strip presents no chemical hazard of its own: its silver is metallic and held in hardened gelatin, and the soluble silver–thiosulfate complexes leave in the wash. None of those absences is a statement about the substances in general — each is about this operation, at these concentrations and quantities, which is what a hazard assessment is.
Hazards
Section titled “Hazards”| Hazard | Where it arises | Control |
|---|---|---|
| Skin sensitisation from metol and hydroquinone | Eleven cylinders and three tanks poured, agitated and emptied over three hours | Single-use nitrile gloves throughout, changed if one is splashed inside. The glove page has the compatibility argument |
| Splashes to the eye, and working in darkness beside open vessels | Lifting strips once a minute with the light off | Eye protection from the first pour to the last emptied vessel; cylinders filled with headroom; the bench laid out in the light and not moved; one hand wet and one dry |
| Alkaline solution standing unattended for an hour | The two floor-arm cells | Labelled on the vessel before filling, in a tray that holds the whole volume, outside every working reach, never topped up |
| Spent fixer, acidic and silver-bearing | Emptying the fixing tray after fourteen pieces of film | Its own labelled container to the silver stream, never combined with developer waste |
| Loss of an exposure batch that cannot be replaced | Eleven strips and one roll exposed in single sittings | The tin stays closed except during loading; strips notched by touch before any light goes on |
Required PPE
Section titled “Required PPE”Single-use nitrile gloves for every wet operation, donned and removed per the PPE SOP, because metol and hydroquinone are skin sensitisers and a sensitisation, once acquired, does not go away. Eye protection from the first pour to the last emptied vessel. And clean dry hands for the film, which protects the film rather than you: this page measures density differences of a few hundredths, and a fingerprint on a uniform strip is a difference of exactly that size in exactly the place you will read.
Respiratory protection is not among the controls, for a specific rather than a reassuring reason: nothing here is a powder, nothing is heated, and no step produces a vapour or an aerosol. The particulate control belongs to the mixing lab.
Ventilation
Section titled “Ventilation”Ventilation is not the control this page relies on, because nothing here evaporates or aerosolises at 20 °C: the developer is a dilute aqueous solution of an aminophenol salt, a phenol, a sulfite and a borate, and the fixer a thiosulfate at working strength. The room needs ordinary comfort ventilation for three hours, part of it in the dark, which Part XVI’s darkroom build provides through a light-tight vent. If the fixer is an acidic rapid type and the room is small, ventilate as the fixer mixing SOP sets out.
Materials
Section titled “Materials”| Item | Quantity | Note |
|---|---|---|
| Test strips from one exposure batch | 11 | 135 mm, one emulsion batch: seven wedge, four uniform |
| A 36-exposure roll of the same film, exposed on an even-toned subject | 1 | Cut into three 490 mm lengths in the changing bag |
| Opaque batch tin | 1 | Opened only in darkness or inside the bag |
| Transmission step wedge | 1 | The Stouffer T2115 or equivalent, the same one throughout |
| Card label slips | 14 | The eight-field slip from the test-negative lab. A strip without one is not data |
| Measuring cylinders, 100 mL, dedicated to developer | 7 | One per cylinder cell, standing in the bath |
| One-reel 35 mm tanks with spirals | 3, or 1 used three times | Three puts all the roll lengths in one half hour |
| Water bath, and an overflow tray for the standing cells | 1 each | The bath deep enough for seven cylinders and a tank; the tray big enough for two cylinders’ contents |
| Film clips and a drying line | 14 clips | Still, dust-free air. Do not squeegee anything here |
| Negative sleeves | 8 pockets | Strips in pairs, roll lengths singly |
| Curve-plotting worksheet, lab notebook sheets and the formula version record | 7 + 1 set + 1 | One plotting sheet per wedge strip; the version record names the bottle |
Chemicals
Section titled “Chemicals”Nothing is weighed. Every solution is diluted from a stock or a concentrate, per the SOP for mixing from a stock.
| Chemical | Quantity | Form |
|---|---|---|
| D-76 stock from Part VIII’s lab — metol, hydroquinone, sodium sulfite, borax | 400 mL | Diluted 1+3 or 1+19 immediately before use and discarded after one cell, as Kodak’s J-78 sheet directs |
| Plain water rinse at the cell’s temperature | 3 L, two changes per piece of film | ILFORD note that a water rinse may replace a stop bath but increases the risk of processing marks and stains |
| Rapid fixer, ammonium thiosulfate type | 750 mL at 1+4 | Twice the clearing time, per the clearing-time SOP |
| Wetting agent | 750 mL at 1+200 | ILFOTOL at 5 mL per litre; ILFORD warn that too little or too much leads to uneven drying, which here would be read as a result |
The silver bromide in the emulsion is a reagent rather than a substance you handle, and the bromide it releases is the subject of the experiment: made inside the layer, leaving by diffusion, at a rate agitation decides. p-Aminophenol appears only in Further experiments.
Equipment
Section titled “Equipment”The sensitometer from Part XIV with its certificate and uniformity map, used per the exposure-session SOP and, for the uniform strips, with the wedge lifted off and a short exposure — a mode the build supports and the calibration page did not test. The densitometer from Part XV with its certificate, 2.0 mm aperture and file logging, checked per its SOP. A camera, a changing bag, one thermometer checked per the bench-check SOP, a stopclock, a millimetre rule for the profile positions, and a spreadsheet. None of it is consumed, which is why none of it appears in the table below.
Estimated cost
Section titled “Estimated cost”££ on the planner’s bands, almost all of it capital bought earlier: the two instruments, the wedge, the cylinders, the tanks, the camera and the thermometer. This session spends two cassettes of film, about 400 mL of D-76 stock and about 150 mL of fixer concentrate.
Estimated consumables cost
Section titled “Estimated consumables cost”One run, at eleven strips, one roll and fourteen pieces of film through the fixer:
| Consumed | This session | Sourced price | Cost this session |
|---|---|---|---|
| 35 mm film for the strips | one cassette, cut into eleven 135 mm strips | £6.37–£11.40 per 36-exposure roll | £6.37–£11.40 |
| 35 mm film for the continuous-tone arm | one cassette, exposed whole | £6.37–£11.40 per 36-exposure roll | £6.37–£11.40 |
| D-76 stock, from Part VIII | about 400 mL, diluted and discarded cell by cell | Costed in Part VIII’s mixing lab | — |
| Rapid fixer concentrate | 150 mL, to make 750 mL at 1+4 | £21.05–£25.98 per 1 L, diluted 1+4 for film | £3.16–£3.90 |
| Wetting agent | 3.75 mL, for 750 mL at 1+200 | £28.70 per 1 L, diluted 1+200 | £0.11 |
| Water for rinses and washes | about 12 L | Metered supply; the planner prices no water | — |
| Negative sleeves and card label slips | 8 pockets, 14 slips | None. A named price gap: sleeving that passes the Photographic Activity Test | — |
| Squared paper for the plots and profiles | 8 to 10 sheets | None. The planner carries no stationery line | — |
The priced rows come to £16.01 to £26.81, at the retail ranges read on 5 September 2026 and recorded in the laboratory planner. That is a floor rather than a total: four of the eight rows carry no dated price, so they count as nothing here and are certainly not free, and a priced entry is a dated range to plan against rather than a quotation. Both film rows price a cassette because that is what the price file carries a figure for; a bulk tin is far cheaper per strip and the file has no bulk-tin price, so the cheaper route is the one this table cannot cost.
Waste streams
Section titled “Waste streams”Three, kept apart, labelled per the container SOP and routed per the general waste SOP.
- Spent developer, about 1.75 L of D-76 at 1+3 and 1+19, alkaline, carrying metol, hydroquinone, sulfite, borate and the bromide released by fourteen pieces of film. Kodak’s instruction is that a diluted solution is discarded after its batch, so all of it is waste.
- Spent fixer and the first rinse after it, silver-bearing, to the silver stream.
- Rinse and wash water: the two rinses after each development go with stream 1, the first change after fixing with stream 2, the rest with the general stream.
Keep streams 1 and 2 apart: an alkaline sulfite solution poured into an acidic silver-bearing one spoils the recovery of the silver and wastes both containers.
Alternative route
Section titled “Alternative route”Darkness is needed for two operations: taking strips from the tin and notching them, and the cylinder cells themselves, because strips are lifted from open vessels. The camera roll, the tank cells, fixing, washing, drying, reading and all the arithmetic happen in room light or inside a changing bag.
Route one: the all-tank version, which removes the darkroom entirely. Load every piece of film onto spirals in a changing bag and run every cell in a closed tank. Two things improve — the agitation becomes inversion, the form every manufacturer publishes, so the control stops being a translation, and a closed tank holds temperature better. Three get worse: one cell per tank, so eleven strips become five or six sequential runs across two evenings; the volumes rise, because a one-reel tank needs about 290 mL where a cylinder needs 80, which moves every cell on the agent-floor plot below, so recompute it rather than copying these numbers; and the developer temperature can only be read going in and coming out, so log both and treat the difference as the term the cylinder would have measured.
Route two: ILFORD ORTHO Plus, whose notching and cylinders run under a deep red safelight — though the control’s published time must then be ORTHO Plus’s own. Route three: a windowless room at night, proved rather than assumed with the blackout leak test.
If none is available, the cylinder arm cannot be run — but route one is a complete substitute needing only a changing bag, a tap and patience, so this page has no dead end, only a slower route with a different vessel. The report must name which vessel produced the numbers, because the vessel is part of the result.
Preparation
Section titled “Preparation”About 75 minutes the evening before, plus two exposure sessions. The 180 minutes here is the wet run; the film cannot be read until bone dry, so the densitometer session is a third sitting of about 90 minutes. Every definition below is fixed before a number exists, because a decision made afterwards is a preference.
What the sources actually say about stand development
Section titled “What the sources actually say about stand development”Read this before designing anything, because it decides what the experiment can be.
No manufacturer of a general-purpose film developer in the course’s corpus publishes a stand or minimal-agitation procedure. ILFORD publish an inversion scheme and warn that very short times with some films may lead to uneven processing. Kodak publish one and add that agitation “should always consist of irregular or random movements that will not cause solution currents to flow over the film constantly in any one direction; these currents increase film density along their paths, causing nonuniformity”. Bergger, selling a developer on its definition, require agitation every 15 seconds and list “denser image edges” against inadequate shaking.
Two sheets in the corpus do, and both are for the same single developer. Photographers’ Formulary and Bostick and Sullivan both print Sandy King’s recommendations for Pyrocat-HD: minimal agitation at 10 seconds every three minutes with times about 50 per cent longer, and semi-stand at 1 part A to 1 part B to 200 or 400 parts water, agitated for one minute at the start and 30 seconds at the halfway point, for 40 to 60 minutes. Part VIII’s rule stands: a scheme published for one developer is not evidence about another.
No Tier 1 or Tier 2 source in the corpus publishes measured sensitometric data for stand development, and this is the finding that matters most. King is the practice’s leading specialist advocate and publishes contrast-index charts for a dozen films — stating that they “are based on rotary processing in BTZS type tubes”, which is continuous agitation. Even the advocate’s numbers are not stand numbers. The nearest thing to data is Ed Buffaloe’s account on the Formulary sheet of two rolls of Delta 3200: one printed with “extremely high acutance”, and the other was ruined, because “the bromides released by the intense development in the heavily exposed areas diffused out and caused uneven development in surrounding areas”. An honest report of two rolls is not a measurement, and the sheet does not present it as one.
One genuine measurement touches the practice, and it is a century old and about dilution rather than agitation. Wall reports Wratten and Wainwright’s photometry on rodinal: a plate needing 3 minutes at 1:20 needed 42 minutes at 1:200 in air-free distilled water, 46 in ordinary distilled water and 52 in tap water, against the 30 proportionality predicts. Time does not scale with dilution, the error runs 40 to 70 per cent towards more time, and its size depends on dissolved air — so a stand arm designed from arithmetic comes out under-developed by an amount that depends on your tap.
The mechanism the whole experiment turns on
Section titled “The mechanism the whole experiment turns on”Development consumes developing agent at the film surface and gives back bromide.
Both happen inside the gelatin, so everything the bath does has to cross the layer of solution sitting on the emulsion — the diffusion boundary layer. Transport across it is diffusion, and diffusion obeys Fick’s first law.
J is the flux in moles per square metre per second; D the diffusion coefficient, between about 0.6 and 2 × 10⁻⁹ m² s⁻¹ for small ions at room temperature; dc/dx the concentration gradient across the layer. Agitation changes none of the chemistry in the two equations above: it changes x, and therefore the gradient and the flux. Kodak’s primer of 1928 put it in one sentence, that the speed of development depends chiefly on the rate at which the developer diffuses into the film, and Kodak’s modern process control document says the same thing operationally: agitation maintains uniform solution activity by removing exhausted solution from the emulsion surface and replacing it with fresh.
The boundary layer at the emulsion surface under three regimes
- Bulk developer — the composition on the label: unchanged, and never what the emulsion sees
- The boundary layer — stagnant solution; its thickness is the only thing agitation sets
- Agent profile (solid) — falls towards the surface, and falls furthest where density is high
- Bromide profile (dashed) — rises towards the surface, and rises furthest where density is high
- The loaded layer sinking — denser than bulk, so it flows downwards over whatever is below
One mechanism, three scales, three names
Section titled “One mechanism, three scales, three names”Everything else on this page falls out of that picture by changing only the distance the bromide has to travel before it reaches solution that is being replaced.
- Micrometres — the adjacency effect. Where a heavily exposed region abuts a lightly exposed one, the inhibiting by-products diffuse sideways into the thin side while fresh agent diffuses the other way. Development runs faster at the edge of the dense area, giving a local density maximum (the border effect), and slower at the edge of the thin area, giving a local minimum (the fringe effect). That is the mechanism as the peer-reviewed literature states it, measured with a microdensitometer on traces whose distance axis is in micrometres.
- Millimetres to centimetres — bromide drag. The loaded layer over a dense area holds dissolved bromide and oxidised agent, is denser than the bulk, and sinks, retarding development in whatever it flows over: streaks below dense areas, running with gravity rather than with the picture. The troubleshooting entry has the diagnostic form.
- The whole frame — compensation and local exhaustion. A highlight consumes its local reservoir and slows; a shadow never does. The upper scale flattens and the lower does not — a shoulder manufactured in the tank rather than coated into the film.
One mechanism, two names, decided by scale
- The knife edge — a dense region above a thin one: the geometry both effects need
- Border and fringe — a density maximum on the dense side and a minimum on the thin side, tens of micrometres wide
- The drag streak — the same chemistry carried by gravity, millimetres to centimetres wide
- What the instrument can see — the 2 mm aperture averages the top inset away and measures the bottom one directly
The three regimes, written out as scripts
Section titled “The three regimes, written out as scripts”A regime described as “occasional” is not data. Write these out, tape them to the bench, and record what you actually did against what you meant to do.
In a cylinder — a 135 mm strip standing in 80 mL in a 100 mL measuring cylinder, lifted by its top edge:
- G-C, continuous. Gently up and down for the whole 16 minutes without pause, about one cycle per second. ILFORD’s instruction is that continuous agitation reduces a spiral-tank time by up to 15 per cent, so at equal time this cell is expected to over-develop, by a published amount rather than a guessed one.
- G-I, intermittent — the control. Gently up and down for the first 30 seconds; then lift, drain for two seconds and re-immerse once at the start of each subsequent minute. This is Part IX’s script, and it is the course’s translation of ILFORD’s four inversions per minute into a vessel that cannot be inverted. Call it what it is in the report.
- G-S, semi-stand. Gently up and down for the first 30 seconds, then nothing except a single 10-second agitation at 8:00 — the halfway interval King’s published semi-stand scheme uses, transplanted as a shape and not as a specification, because it was published for a different developer.
In a tank — a 490 mm length of roll film on a spiral in 290 mL. R-H, the most agitation any maker publishes: Kodak’s small-tank procedure of 5 to 7 inversion cycles in 5 seconds at 30-second intervals. R-I, the control in its published form: ILFORD’s four inversions during the first 10 seconds and four again during the first 10 seconds of each further minute, tapping the tank afterwards to dislodge bubbles. R-S, semi-stand: ILFORD’s opening four inversions, then nothing until a single four-inversion sequence at 8:00.
The developing-agent floor
Section titled “The developing-agent floor”This part of the design has nothing to do with agitation, and it is the reason half the arguments about stand development are unresolvable.
Kodak state it plainly for XTOL, and the reasoning is general even though the number is not: the minimum amount of diluted developer needed to cover the film may not contain enough active ingredients to develop the film fully in the recommended time, so they recommend at least 100 mL of full-strength developer for each 135-36 or 120 roll, which they give as 80 square inches or 516 cm². Their D-76 sheet says the same in a different currency: one roll in 473 mL of 1+1, which is 236 mL of stock, or in 237 mL with the time increased by 10 per cent, which is 118 mL. Convert everything into one quantity before comparing anything:
Vbath is the working-solution volume in millilitres, n the parts of water per part of stock, and A the film area in square inches. A 135 × 35 mm strip is 47.25 cm², or 7.32 in² — a little over a ninth of a roll.
Developing agent per unit film area, against dilution, for the two vessels this session uses
- One 135 mm strip in 80 mL — this session's cylinder cells
- One whole 36-exposure roll in a 290 mL tank
Show the numbers behind this plot
| Series | Parts of water per part of D-76 stock (the n in 1+n) | mL of stock developer per 80 in² of film |
|---|---|---|
| One 135 mm strip in 80 mL — this session's cylinder cells | 3.00 | 218.40 |
| One 135 mm strip in 80 mL — this session's cylinder cells | 9.00 | 87.40 |
| One 135 mm strip in 80 mL — this session's cylinder cells | 19.00 | 43.70 |
| One 135 mm strip in 80 mL — this session's cylinder cells | 39.00 | 21.80 |
| One whole 36-exposure roll in a 290 mL tank | 1.00 | 145.00 |
| One whole 36-exposure roll in a 290 mL tank | 3.00 | 72.50 |
| One whole 36-exposure roll in a 290 mL tank | 9.00 | 29.00 |
| One whole 36-exposure roll in a 290 mL tank | 19.00 | 14.50 |
| One whole 36-exposure roll in a 290 mL tank | 39.00 | 7.25 |
Three things follow, and together they justify the floor arm. The fair test is deliberately far above the floor: both arms sit near 215 mL of stock per 80 in², close to Kodak’s generous D-76 figure of 236, so no regime is short of agent and the comparison is about transport rather than supply. The floor arm is deliberately below it: two cells at 1+19 in the same 80 mL sit at 44 mL per 80 in², under half Kodak’s XTOL minimum, so if a curve compresses at the top there, exhaustion is a live explanation.
And the ordinary home stand procedure is below the floor before it starts. A whole roll in a 290 mL tank is at 73 mL per 80 in² at 1+3 and 29 at 1+9, and a stand procedure would dilute further still. That is the trap: the worker sees a negative thin in the highlights, calls it compensation, and it is starvation. The two do not respond the same way — compensation is undone by more agitation, starvation only by more agent.
Compensation, defined before there is any data
Section titled “Compensation, defined before there is any data”“Compensating” is a word this course will not use qualitatively on a page with a densitometer on the bench. The definition below is the course’s own construction, stated here because no page before this one needed a number for it.
Pure Silver compensation ratio (R). Plot diffuse density against log exposure. Take base-plus-fog D₀ from the strip’s own masked patch. Locate the anchor m, the point where the curve reaches D₀ + 0.10 — the same point the course’s speed criterion uses. Then
Four notes. The intervals are 1.00 log H each, three and a third stops, and the pair spans 2.00 — the same span as the contrast-index chord, so a curve that supports one supports the other. On a T2115 that is thirteen and a third steps against a usable scale of about sixteen. Interpolate linearly between the two steps straddling each exposure, and say so.
R is compared between cells, not against unity. A negative whose toe is still rising through the first interval usually gives R above 1, because the second interval is straight line where the first is part toe. Compensation shows as R falling relative to the control, and a cell whose R drops below 1 has an upper scale genuinely flatter than its lower.
R is nearly immune to the error everyone worries about: both gradients are differences of densities, so a constant offset in the whole scale cancels in each, and again in the ratio. And R is not a measure of sharpness and must never be reported as one — it describes the shape of the characteristic curve, which is a whole-frame quantity, while the edge effects live three orders of magnitude below the aperture.
Evenness, defined before there is any data
Section titled “Evenness, defined before there is any data”Nothing earlier in the course has measured this, and it is the quantity that makes the experiment worth running.
The evenness measurement. On a uniformly exposed strip, read density at 20 positions at 5 mm pitch along the strip’s length, starting 20 mm from the head, with the aperture centred across the strip’s width. Report the standard deviation s of the twenty readings, the range, and the profile itself. Then subtract the control cell’s profile position by position and report the standard deviation of the difference profile, which is the figure the cells are compared on.
Why each choice. Twenty positions at 5 mm pitch, because the aperture is 2.0 mm so the readings do not overlap, the run covers 95 mm of a 135 mm strip, and twenty is enough for s to mean something. A standard deviation rather than a range, because a range grows with the number of readings and is decided by one outlier, which on film is as likely to be a dust speck as a fault; quote the range too, but compare on s.
The difference profile is the headline, because every strip carries the same systematic pattern before development starts. The sensitometer’s field is a shallow dome, computed on the enclosure page as falling 0.053 log H at the ends of the 127 mm field — at a gradient of 0.6, 0.032 D of built-in curvature, the same size as the effect being looked for. Subtracting the control’s profile removes the dome, the coating variation and any common systematic in one operation, with no modelling, at a cost of √2 in noise. The control’s own s is then the floor: it contains the source’s non-uniformity, the coating, the densitometer’s replaced-sample repeatability and your hand on the rule, and no cell can be reported as more even than that.
Two profiles with the same standard deviation and different diagnoses
- A control-like profile: small scatter, no shape
- A resolved feature: a broad trough, as a drag streak reads
- Unresolved fine structure: scatter with no shape
Where an even exposure actually comes from
Section titled “Where an even exposure actually comes from”The uniform strips need a source flat enough that its own structure does not swamp what is being measured. Three candidates are on the course’s bench, and only one of them qualifies.
The sensitometer with the wedge lifted off, which is the choice. Same box, lamp, fences and firmware,
and a field whose non-uniformity is already a number in your certificate. The one change is the exposure:
the box is dimensioned to expose through up to 3.05 D of wedge, so without it the film sees up to a
thousand times more light. The design exposure is 500 ms; the uniform strips will want something near 5
to 20 ms, and the firmware reports measured_us for every flash, so what was delivered is logged rather
than assumed. Bracket it first — four scrap strips at 4, 8, 16 and 32 ms developed together in the
control regime — and take the one landing about 0.9 above base, where the gradient is high and a small
development difference shows most.
The enlarger, which is the fallback if your timer cannot deliver a repeatable few milliseconds: the head with no negative in the carrier, stopped down, strips on the baseboard in darkness. Its uniformity is mapped in density units on the material by the commissioning page, whose acceptance criterion is opposite corners within 0.04 D; the cost is a second instrument with a second calibration.
The pinhole camera, which is disqualified, because illumination on its film plane falls off as cos⁴ of the field angle and Part VI computes 3.7 stops of corner falloff for the course’s reference camera. A source whose own field varies by four stops cannot measure evenness to a few hundredths of a density.
The camera roll is not a measurement instrument either. It is exposed through a lens with its own falloff, on a real subject, and it is here because mottle and drag show themselves in continuous tone and on perforated film. It yields photographs of defects and a difference profile, not an absolute evenness figure — and the difference profile is legitimate because the lens’s falloff is identical in every frame and cancels when one regime’s frame is subtracted from another’s.
The run order and the table drawn before the data
Section titled “The run order and the table drawn before the data”Eight cells, in an order drawn from a hat and written down beforehand, so that a drift in the bath, the lamp’s age or your own attention cannot line up with the variable. Two constraints bind the draw: the continuous cell occupies both hands for a quarter of an hour and takes a batch to itself, and the floor cells occupy an hour, so they start first with everything else run underneath them.
| Cell | Vessel | Dilution | Volume | Time | Regime | Strips |
|---|---|---|---|---|---|---|
| G-C | 100 mL cylinder | 1+3 | 80 mL | 16:00 | Continuous | wedge, uniform |
| G-I | 100 mL cylinder | 1+3 | 80 mL | 16:00 | Intermittent (control) | wedge, uniform, wedge replicate |
| G-S | 100 mL cylinder | 1+3 | 80 mL | 16:00 | Single agitation at 8:00 | wedge, uniform |
| F-16 | 100 mL cylinder | 1+19 | 80 mL | 16:00 | Single agitation at 8:00 | wedge, uniform |
| F-60 | 100 mL cylinder | 1+19 | 80 mL | 60:00 | Single agitation at 30:00 | wedge, uniform |
| R-H | one-reel tank | 1+3 | 290 mL | 16:00 | Kodak’s 30-second scheme | 490 mm roll length |
| R-I | one-reel tank | 1+3 | 290 mL | 16:00 | ILFORD’s per-minute scheme | 490 mm roll length |
| R-S | one-reel tank | 1+3 | 290 mL | 16:00 | Single sequence at 8:00 | 490 mm roll length |
Read the table for what it admits. G-C, G-I and G-S differ in one thing. F-16 differs from G-S in one, the dilution, and therefore in agent per unit area. F-60 differs from F-16 in one, the time — and so from G-S in two, which is why it is never quoted as an agitation result. R-H, R-I and R-S differ in one and share the control condition, so the arms are tied together at G-I and R-I and nowhere else.
Draw the empty data table now, a row per piece of film and a column per quantity, with a dash in every cell: a table drawn afterwards has its interesting rows already chosen.
Procedure
Section titled “Procedure”After the exposure batches: preflight, 30 minutes; allocating the film in darkness, 15; the floor cells started, 5; the cylinder arm, 45; the tank arm, 55 nested inside the floor cells; wash, dry and file, 30; and the reading session on another evening, 90.
Stage 0 — The two exposure batches
Section titled “Stage 0 — The two exposure batches”- The strips, in one sitting, from one lamp at one distance, per
the exposure-session SOP, logging the minutes since switch-on
against each: seven wedge strips plus two spares through the wedge, then, with the wedge and cover glass
lifted off and the mask still in place, four uniform strips plus two spares at the single bracketed
exposure chosen in Preparation, logging
measured_usfor each. Every strip carries the masked patch the test-negative lab specified, and the uniform strips are notched on the opposite edge. - The continuous-tone roll. Expose a whole cassette on one even-toned subject at one meter setting in one session — a matt wall lit obliquely, or an overcast sky well away from the sun — filling the frame every time, aiming for the same mid-grey as the uniform strips. Do not bracket.
- Store everything in the closed tin and develop as soon as the calendar allows: ILFORD’s instruction is to process exposed film as soon as practical, and the course has no figure for what months of keeping do to a latent image.
Stage 1 — Preflight
Section titled “Stage 1 — Preflight”- Check the thermometer per the bench-check SOP and write the offset down. Bring the bath to 20 °C and give it twenty minutes to settle with the empty vessels in it. Kodak state that a developer temperature varying by more than 0.3 °C affects process control and image quality; ILFORD ask for all solutions within 1 °C. If your bath cannot hold that, write the figure it can hold and carry it through the analysis.
- Lay the bench out in the light in the order it will be used and find the light switch by touch. Make up the fixer at 1+4 and the wetting agent at 1+200, bring the rinse water to 20 °C, and measure the clearing time on a scrap per the clearing-time SOP; fixing time is twice it.
- Label the two floor cylinders on the vessel and set them in the overflow tray, in their final place, before anything is poured. They stand there for an hour.
Stage 2 — Allocate the film, in darkness
Section titled “Stage 2 — Allocate the film, in darkness”- With the light off, nip each strip along its top edge to its cell — one nip for G-C through five for F-60 — with the uniform strips nipped on the lower edge so they cannot be confused by touch. Close the tin before the light goes on.
- In the changing bag, cut the continuous-tone roll into three 490 mm lengths, notch each and load each onto its own spiral. The length is not arbitrary: at 290 mL of 1+3 it puts the tank cells at the same agent per unit area as the cylinder cells.
Stage 3 — The floor cells, started first
Section titled “Stage 3 — The floor cells, started first”- Dilute 160 mL at 1+19 — 8 mL of stock into 152 mL of water — split it between the two F-16 cylinders and start the clock; then a second 160 mL for F-60, a minute later. Both get 30 seconds of gentle movement and nothing until their single agitation.
- Record the volume, the dilution and the millilitres of stock as a number, not a ratio. Four millilitres of stock per strip is the point of these cells, and a ratio hides it.
Stage 4 — The cylinder arm
Section titled “Stage 4 — The cylinder arm”One evening, eight cells: an example draw
- Batch one: G-S and G-I. Dilute one 160 mL jug of 1+3 for G-S and one 240 mL jug for G-I, whose control cell has three strips in three cylinders, pouring each cylinder immediately before its cell begins. The semi-stand cell needs no hands, so start it a minute early and leave it.
- Log the developer temperature at the start, middle and end of every cell, with the thermometer in the cylinder rather than the bath. Then lift each strip at its own mark, rinse for one minute in two changes of plain water at 20 °C, and fix for twice the clearing time.
- Batch two: G-C. Fresh 160 mL of 1+3, two cylinders, hands on both for the full 16 minutes. Kodak’s instruction is that a diluted solution is discarded after its batch, and a bath that has already developed a strip is a different bath — the subject of the exhaustion experiment, which must not leak into this one.
Stage 5 — The tank arm, nested inside the floor cells
Section titled “Stage 5 — The tank arm, nested inside the floor cells”- Fill each tank with 290 mL of freshly diluted 1+3 and start it on its written scheme. With one tank, run the three in sequence and record each clock time, because the bottle is an hour older by the third.
- Tap the tank after every agitation sequence, as both makers instruct, to dislodge bubbles: a bubble on the emulsion during a stand cell stops all exchange of developer at that spot, and King records that the resulting circular artefact ruins the frame. Rinse and fix as for the strips, and do not open a tank early to look.
Stage 6 — Wash, dry, label and file
Section titled “Stage 6 — Wash, dry, label and file”- Wash everything for 5 to 10 minutes in running water within 5 °C of the process temperature, or use ILFORD’s spiral-tank sequence of five, ten and twenty inversions in three changes. Final-rinse in wetting agent at 1+200 and hang in still, dust-free air. Do not squeegee anything: ILFORD warn that too little or too much wetting agent leads to uneven drying, and here an uneven dry reads as a result.
- Complete a card slip for every piece of film and record the developer stock’s formula version code, issued per the versioning SOP on the formula version record. A result tied to “the D-76 on the shelf” is a result nobody can check.
- When everything is bone dry — not before — sleeve and file it. Do not read anything tonight: film that is not quite dry reads high.
Stage 7 — The reading session, on another evening
Section titled “Stage 7 — The reading session, on another evening”- Run the densitometer’s warm-up and daily check per its SOP, recording the calibration step before the first piece of film and after every third.
- Read all twenty-one steps and the masked patch of every wedge strip, in a separately drawn reading order, logging to a file. Read one strip twice, lifting it off the stage and replacing it between readings: that replaced-sample spread is this session’s Type A term.
- Read the twenty profile positions on every uniform strip, marking the sleeve so the positions match across strips to better than a millimetre, then five transverse positions at the midpoint as a check for a cross-strip tilt.
- Read a twelve-position profile across one frame of each roll length, in the same place in the frame on all three, from one perforated edge to the other — the direction gravity ran while the film was coiled.
Expected observations
Section titled “Expected observations”What follows is what the design expects, and what would show the expectation wrong. It is not a prediction of your numbers: a page that supplies those has run the experiment for you.
In the trays. The three fair-test strips should look more alike than the two floor strips do, and the 1+19 cell at 16 minutes should come back visibly thin — not a spoiled strip but the cell doing its job. Whether an hour is enough at 1+19 is genuinely open, for the reason Wratten and Wainwright measured, and finding out is part of what the 60-minute cell is for.
Where to look for a defect. On the cylinder strips, drag runs along the strip, because the strip stood vertically. On the roll lengths gravity ran across the film, because it was coiled, so look for streaks running inwards from the perforated edges and for regular marks at the perforation pitch — measure that pitch on your own film rather than assuming it, because the course has not verified the standard figure. Kodak’s troubleshooting table places streaks of non-uniform density against excessive or uneven agitation and mottle against inadequate agitation, so both ends of the ladder are expected to fail, and differently; the surge-marks entry is explicit that the course has found no manufacturer publication naming the sprocket-hole mark specifically.
Five things that would falsify the design rather than the chemistry. The control cell’s two wedge strips disagreeing by more than the resolution sentence allows. The control’s uniform profile having a standard deviation as large as the differences you hoped to see. The 1+19 cells clearing the top of the wedge, which would mean the dilution did not cross the floor. Any cell’s temperature log leaving the stage 1 tolerance. And R-I disagreeing with G-I by more than the between-strip spread, which would say the two vessels are not running the same process and the arms cannot be tied together at all.
What is happening chemically
Section titled “What is happening chemically”Bromide does two jobs, and the experiment has to keep them apart
Section titled “Bromide does two jobs, and the experiment has to keep them apart”The transport account is in the Preparation, above; what it leaves to be said is what the bromide does once the boundary layer has held onto it. Chemically it is a restrainer, acting through the common-ion effect at the crystal surface:
That is a concentration effect and it acts wherever the bromide is. Well agitated, the released bromide disperses into 80 or 290 mL and the change is negligible; in a still bath it stays within a fraction of a millimetre of where it was made, and the local concentration can be large. The chemistry Part IX’s bromide arm studied by adding potassium bromide to a whole bath is here generated selectively by the image, in proportion to local density. A still bath restrains itself in exactly the places developing hardest — compensation in one sentence, and the reason Kodak can say that agitation affects the rate of development particularly in high-density areas.
The second job is physical, and nothing chemical happens during it: the loaded layer is denser than the bulk and sinks, arriving somewhere it was not made and restraining development there, giving a streak whose shape follows gravity and the film’s geometry rather than the picture.
Why the compensating shape and the starved shape look alike
Section titled “Why the compensating shape and the starved shape look alike”Compensation flattens the upper scale because the highlights run their local reservoir down faster than diffusion refills it. Starvation flattens it because the whole bath never held enough agent to develop them out. The shapes are similar and the causes are not, and the discriminator is built into the design: compensation is a function of transport, starvation a function of supply. Two cells at one dilution and volume differing only in agitation isolate the first; two cells at one agitation and volume differing only in dilution isolate the second. That is the whole reason the floor arm exists, and it is why a claim about stand development made without a stated volume of stock per unit area is not checkable.
A third mechanism sits at the same end of the curve and must not be attributed to either: in a very dilute bath standing for an hour with a large surface open to the room, the agent is also consumed by air. Aerial oxidation is why Wall could already call stand development uneconomical in 1924. The floor cells stand longest and are the most exposed solution in the session, so oxidation is a real term in them that nothing here separates from starvation — say so rather than choosing between them.
Data to record
Section titled “Data to record”Session header, as the design lesson sets out: date, operator, room temperature, the developer’s formula version code, both instrument certificates by date, the thermometer’s offset, the wedge’s serial and whether it is calibrated, and the drawn run and reading orders.
Per cell, before any film enters it:
| Field | Why |
|---|---|
| Vessel and its volume | The vessel is part of the result |
| Dilution as 1+n, and millilitres of stock | The ratio hides the quantity that matters |
| Film area, and mL of stock per 80 in² | The floor comparison is impossible without it |
| Agitation regime, as intervals and durations | “Occasional” is not data |
| Time, to the second, as run rather than as planned | |
| Temperature at start, middle and end, in the developer | Decides whether the cell is comparable |
| Clock time the cell started | The bottle ages during the session |
Per wedge strip: all twenty-one step densities, the masked patch, the replicate readings, and the derived base plus fog, contrast index, relative speed point, maximum density, Glow, Ghigh and R, each with its uncertainty.
Per uniform strip: twenty profile densities with their positions in millimetres, the five transverse readings, the mean, the range, s, and s of the difference profile against the control.
Per roll length: the twelve-position transverse profile of the nominated frame, and a photograph of every defect, taken on a light box with a rule in the frame before the film is sleeved and indexed to the notebook page. A defect described in words is an anecdote; one photographed with a scale beside it is evidence.
Analysis
Section titled “Analysis”1. The resolution sentence, restated for this page’s five quantities
Section titled “1. The resolution sentence, restated for this page’s five quantities”Before any comparison, write down what your instrument can report, by the propagation method the curve-families page sets out. Five quantities, five different answers from the same readings:
| Quantity | How its uncertainty is got | Rough size at u(D) = 0.01 |
|---|---|---|
| Maximum density | Read directly | 0.01; a difference between cells, 0.014 |
| Contrast index | Perturb the density table and re-solve the construction a few thousand times | About the same as one reading |
| Relative speed point | The same perturbation; the multiplier is one over your own toe gradient | Three to four times worse than a density |
| Compensation ratio R | The arithmetic in Preparation | 0.044, so a difference below about 0.12 is not a result |
| Evenness s | Sampling uncertainty of a standard deviation from n = 20 | About 16 per cent of s; a ratio of two must exceed about 1.5 |
The last row is the one nobody expects. The relative standard error of a standard deviation from n readings is 1 ÷ √(2(n − 1)), which at n = 20 is 16 per cent; a ratio of two carries 23 per cent, so two cells’ evenness figures must differ by a factor of about 1.5 before the difference is a two-sigma result. Resolving a factor of 1.2 would need about forty positions on a strip 135 mm long.
2. The three fair-test cells
Section titled “2. The three fair-test cells”Plot the three curves on one field, tabulate the five quantities with their uncertainties, and answer three questions in this order, writing each answer down before moving on.
Did the contrast move, and by how much in units of time rather than contrast? ILFORD state that continuous agitation reduces a spiral-tank time by up to 15 per cent, so at equal time that cell is expected to over-develop by a published amount. Convert each cell’s contrast index into an equivalent development time on the curve you built in the time series and report the regimes as time differences. That is the fair currency: this regime at 16 minutes did what the control would have done at 14, or at 19, and a photographer can act on that.
Did the speed point move, and does the contrast change alone account for it? A change in gradient moves the point at D₀ + 0.10 even when sensitivity is untouched, so read the shift, convert it to stops and subtract the shift the contrast change predicts. A speed difference is only a speed difference if it survives that subtraction — which is how King’s claim of increased emulsion speed under minimal agitation gets tested rather than repeated.
Did R fall as agitation fell? Report ΔR against the control with its uncertainty and against the 0.12 band. If the ordering is right but the differences sit inside the band, the honest conclusion is that the effect on this film in this developer at this dilution is smaller than the apparatus can resolve — a result, and a more useful one than a rank order asserted without an error bar.
3. Evenness, and the two ways it fails
Section titled “3. Evenness, and the two ways it fails”Plot all four uniform-strip profiles on one field with the control’s ±2u band on it, then the three difference profiles against the control on a second. Report per cell: the mean, the range, s, s of the difference profile, and one sentence saying whether the profile has a shape.
That sentence is the diagnosis and no number supplies it. A shaped difference profile — a trough, a ramp, a step — is a resolved feature, and the film should be inspected in transmitted light where the profile says it is. A shapeless one with a large s is unresolved structure below the aperture, which is mottle by another name. A shapeless one with a small s is a cell that developed evenly, which is a real finding.
Then check the transverse five. A consistent tilt across the width in every cell including the control belongs to the source or the vessel rather than the regime, and goes in the certificate, not the result.
4. The floor arm: separating agitation from starvation
Section titled “4. The floor arm: separating agitation from starvation”Compare F-16 against G-S first: same regime, vessel, volume, time and temperature, differing only in dilution and so in stock per unit area, by a factor of five. Whatever separates them is a difference in supply and not in transport.
Then compare F-60 against F-16, which differ only in time, and say whether the extra 44 minutes brought the curve back up. If it did, the 16-minute cell was time-starved rather than agent-starved; if it did not — if the upper scale stayed depressed while the lower rose — the bath ran out of what it had, which is the floor doing what Kodak’s sheet says it does.
Finally, put F-60 beside G-S and refuse to draw an agitation conclusion from the pair, which differs in two things. Its use is in the verdict: it is the session’s closest cell to what a photographer actually does when stand-developing a roll, and its position on the agent-floor plot is the sentence to quote whenever somebody reports a compensating negative without saying how much developer was in the tank.
5. The roll-film arm
Section titled “5. The roll-film arm”The tank cells produce three things: a tie to the cylinder arm through R-I against G-I, whose difference in contrast index and in R is the vessel term for the session; a difference profile across a frame for R-H and R-S against R-I, in which the lens’s falloff cancels because it is the same in every frame; and the defects, which are why the arm exists.
Score each defect as a diagnostic atlas does rather than as a forum post does — what you see, where it sits relative to the film’s geometry, its scale in millimetres — then look it up against bromide drag, surge marks, mottled development and uneven density across a strip, and say which entry it matches and which it does not.
6. The verdict, and the form it has to take
Section titled “6. The verdict, and the form it has to take”Four paragraphs, in this order, keeping your number and the published number visibly apart in every one.
What this session measured. Five quantities for three regimes, on one film in one developer at one dilution, time and temperature, in one vessel, by one worker, on one evening, with an instrument whose uncertainty is stated. Every number is the course’s own under the course’s own criteria.
What the sources contain, restated in your own words because it bounds the conclusion: no datasheet for a general-purpose developer publishes a reduced-agitation scheme, the two that do are for one developer nobody may transplant from, and no Tier 1 or Tier 2 source publishes measured sensitometric data for stand development at all.
What the data supports and what it does not. Name each hypothesis and say whether the difference exceeded the resolution; then name what the session could not test — sharpness, because the aperture is three orders of magnitude too coarse; reproducibility, because n is one per regime; generality, because it is one film and one bath.
The cost side. Stand development takes an hour instead of sixteen minutes; leaves a large volume of dilute developer standing and oxidising, which Wall called uneconomical in 1924; and is irreproducible enough that its own advocate advises two back-up shots of any important scene. It also produces a curve shape available a second way: ILFORD publish 8 minutes at stock and 16 minutes at 1+3 as giving the same average contrast on the same film — a fourfold dilution and a doubled time, with normal agitation and no evenness penalty. If what you want is compensation, the sourced route is dilution and time. If what you want is the edge effect, this apparatus cannot tell you whether you got it.
What cannot be concluded from this session
Section titled “What cannot be concluded from this session”- That stand development is better or worse than intermittent agitation. The session measures five quantities, and one of the things photographers care about most is not among them.
- That any result here transfers to another film, developer, dilution, vessel, water supply or worker. Each of those is a variable this session held fixed.
- That an observed sharpness difference is real: it was not measured and cannot be, with this instrument.
- That the floor cells’ behaviour is starvation rather than oxidation. Both were present and the design cannot separate them.
- That an absent defect will stay absent. One roll per regime is one sample, and King’s own account is that the practice ruins some frames some of the time — a rate no session this size can estimate.
Troubleshooting
Section titled “Troubleshooting”| What you see | Likely cause | What to do |
|---|---|---|
| Streaks running down from the dense steps of the semi-stand strip | Bromide drag, the published failure mode of reduced agitation | A result, not a fault. Photograph it with a rule, read a profile through it at 2 mm pitch, and report it |
| The continuous strip is no denser than the control | The agitation was not actually continuous, or the effect is below your resolution | ILFORD’s figure is up to 15 per cent of development time, roughly one wedge step. If your resolution is one step, this comparison was always marginal and the report should say so |
| The control’s two wedge strips disagree by more than the budget allows | Session repeatability is worse than assumed | Nothing here can be reported to a tighter band than that spread. Widen the resolution sentence and re-read every conclusion against it |
| The 1+19 cells are so thin the toe cannot be found | The floor arm crossed further than intended | Read what you can, report those cells as not analysable for R, and repeat the arm at 1+9 rather than interpolating |
| A circular blotch a few millimetres across on a stand cell | An air bubble, which King names as the characteristic ruin of the method | Record it and consider a pre-soak next time — but ILFORD advise against a pre-rinse, so say which advice you followed |
| Regular marks near the perforated edges of a roll length | Solution driven through the perforations | Measure the pitch. If it matches the perforations it is a surge mark, from the over-agitated end of the ladder |
| A gradient along the length of every strip, control included | The sensitometer’s field, a computed 0.053 log H dome | This is why the difference profile exists. Subtract the control and carry on |
Clean-up
Section titled “Clean-up”Empty each developer cylinder into the developer waste container as its cell finishes, rather than leaving a standing cylinder for the end of a long session. Rinse cylinders, tanks and spirals in warm water and stand them to dry — dedicated developer glassware never meets fixer, per the general waste SOP. Wipe the bench, dry the thermometer and put the wedge back in its sleeve, because a fingerprint on a step wedge is a systematic error in every session that follows. Cap the stock bottle full and record on the version record how much was drawn.
Storage
Section titled “Storage”Sleeve the strips in pairs and the roll lengths singly, each with its card slip, in the same sequence as the data file, and file them with the Part IX agitation strips: the two sets are the only between-session replicate the course has for this comparison. Keep the defect photographs with their notebook page. Store the D-76 stock per the storage rotation SOP, tightly capped and full; ILFORD’s instruction for their own diluted developers is that 1+1 and 1+3 solutions are not kept more than 24 hours, so every dilution here is made and discarded on the day.
Disposal considerations
Section titled “Disposal considerations”Three streams, described as chemistry and general practice; what you may actually do with each is set by your local regulations and your water authority, and this page does not override them.
Spent developer is alkaline and carries metol, hydroquinone, sulfite, borate and the bromide released by fourteen pieces of film — a reducing solution with an oxygen demand, containing a boron compound, which is the part most likely to attract a local restriction. Spent fixer and the first rinse after it carry dissolved silver–thiosulfate complexes and go to the silver stream, which exists both because silver is a regulated discharge and because it is recoverable. The remaining rinse and wash water is dilute enough to go with the general stream in most places, and “most places” is doing real work in that sentence. The disposal page has the framework; your authority has the rule.
Questions
Section titled “Questions”Check your understanding
Further experiments
Section titled “Further experiments”The Rodinal-type stand run, done honestly. The practice is done overwhelmingly in a p-aminophenol developer at dilutions like 1+100, and the corpus contains no published time for that pairing at all — ILFORD’s FP4 Plus sheet stops at 1+50. Run it as a declared exploration rather than an experiment: one roll, one dilution, one time, the concentrate per 80 in² recorded, the result reported as a single observation. One safety point from the formulary entry is not optional: the concentrate is strongly alkaline in the bottle even though the working solution is dilute, so handle it with eye protection and add it to water rather than the other way about.
The volume arm this session did not run. Hold the dilution at 1+19 and the regime at semi-stand and change only the volume, 80 mL against 240 mL in a larger cylinder. That crosses the agent floor with one variable instead of two, and costs two strips.
The pre-soak question, where two published sources disagree. King strongly recommends a five-minute pre-soak before any minimal or stand development, to prevent the bubbles that ruin frames; ILFORD state that a pre-rinse is not recommended because it can lead to uneven processing. Run one pair of semi-stand cells with and without, and report which failure you got.
The between-session replicate. Read the archived Part IX agitation strips on today’s densitometer, alongside these: same developer, dilution and time, under three scripts differing only in the semi-stand cell’s timing. The spread between the two sessions is the closest thing this course has to a reproducibility figure.
Sources for this page
18 cited · checked 2026-09-06
- 01FP4 Plus Technical InformationHARMAN technology Limited (ILFORD Photo), 2018§ Development times - Kodak D-76 at stock 6, 8 and 9 minutes, at 1+1 9, 11 and 15 and at 1+3 14, 16 and 20 minutes for meter settings EI 50, 125 and 200 in a spiral tank at 20 degrees C, and Rodinal at 1+25 9 minutes and at 1+50 15 minutes at EI 125; the statement that the times produce negatives of average contrast suitable for printing in all enlargers and are intended as a guide that may be altered if a different result is needed; the instruction that where continuous agitation is used for manual processing the times are reduced by up to 15 per cent, and that a pre-rinse is not recommended because it can lead to uneven processing; the Agitation section, which recommends intermittent agitation for spiral and deep tanks and specifies inverting the tank four times during the first 10 seconds and four times again during the first 10 seconds of each further minute, with continuous agitation recommended for sheet film in dishes; the instruction to keep all process solutions within 5 degrees C of the developer; the spiral-tank wash of five, ten and twenty inversions in three changes of water; ILFOTOL at 5 mL per litre with the note that too little or too much wetting agent can lead to uneven drying; and the instruction to handle the film in total darknessilfordphoto.com/amfile/file/download/file/1919/product/690tier 1, primary2026-09-06
- 02KODAK Developer D-76, technical data sheet J-78Kodak Alaris Inc., 2017§ Agitation - the statement that agitation helps remove the by-products of development from the surface of the film so that fresh developer can act on the exposed silver halide, that agitation affects the rate of development particularly in high-density areas, and the instruction that agitation should always consist of irregular or random movements that will not cause solution currents to flow over the film constantly in any one direction because these currents increase film density along their paths and cause nonuniformity; the small-tank procedure; Development Times - the statement that the tables are starting-point recommendations and that a test should be run for critical applications; the instruction that D-76 diluted 1:1 is diluted just before use and neither reused nor replenished; the volume statement that one 135-36 roll, given as 80 square inches, is developed in 473 mL of the diluted solution, with a 10 per cent time increase where 237 mL is used instead; and the note that tank development times shorter than 5 minutes may produce poor uniformitybusiness.kodakmoments.com/sites/default/files/files/resources/j78.pdftier 1, primary2026-09-06
- 03KODAK PROFESSIONAL XTOL Developer, Technical Data / Chemical, J-109Kodak Alaris Inc., 2018§ Diluting the developer - the statement that the minimum amount of diluted developer needed to cover the film may not contain enough active ingredients to develop the film fully in the recommended time, and the recommendation to start with at least 100 mL of full-strength developer for each 135-36 or 120 roll, or the equivalent of 80 square inches (516 square centimetres), with the worked example that four rolls at 1:1 use at least 800 mL even where the equipment would allow lessbusiness.kodakmoments.com/sites/default/files/wysiwyg/pro/chemistry/J-109_Feb_2018.pdftier 1, primary2026-09-06
- 04PERCEPTOL, ID-11 and MICROPHEN film developers (ILFORD technical information)HARMAN technology Limited, 2024§ Manual processing, spiral tanks - the agitation scheme of four inversions during the first 10 seconds repeated during the first 10 seconds of each subsequent minute, with the tank tapped afterwards to dislodge air bubbles, and the instruction to drain the developer 10 seconds before the end of the development time; the statement that dish processing with continuous agitation reduces the recommended times by about 15 per cent; the warning that very short development times with some films may lead to uneven processing; the instruction that all process solutions be within 1 degree C of the temperature being used; the stated usable range of 20 to 24 degrees C; the statement that the published times should produce negatives of normal contrast, typically around a Gbar of 0.62, and are only a guide; the instruction that 1+1 and 1+3 dilutions are prepared directly before use, are not reused and are not kept more than 24 hours; and the reuse figures of 10 films per litre of ID-11 stock with a 10 per cent time increase per successive film and 250 to 300 mL of solution for one filmilfordphoto.com/amfile/file/download/file/1829/product/550tier 1, primary2026-09-06
- 05Monitoring and Troubleshooting KODAK Black-and-White Film Processes, publication Z-133E, bound with How to Process and Print Black-and-White Film, publication AJ-3Eastman Kodak Company, 2005§ Process control - the statement that agitation is necessary to maintain uniform solution activity by removing exhausted solution from the emulsion surface and replacing it with fresh solution, that it must be uniform throughout the tank, and that a developer temperature varying by more than 0.5 degrees Fahrenheit, that is 0.3 degrees Celsius, affects process control and image quality; Diagnosing and troubleshooting processing problems - streaks of non-uniform density listed against excessive or uneven developer agitation, and mottle, given as areas of non-uniform density, listed against inadequate developer agitation125px.com/docs/techpubs/kodak/z-133-2003_03b.pdftier 1, primary2026-09-06
- 06An Introduction to Pyro Staining Developers, With Special Attention to the Pyrocat-HD FormulaSandy King§ Development of film in tanks, on the second page - Normal Agitation given as continuous agitation for the first 60 seconds then 5 to 10 seconds every 30 to 60 seconds; Minimal Agitation given as continuous agitation for the first 60 seconds then 10 seconds every third minute, with a five-minute pre-soak strongly recommended, three claimed results of great apparent sharpness through maximum adjacency effects, a compensating effect and increased emulsion speed, and development times about 50 per cent longer; and Stand Development, described as a highly specialised method using extremely dilute developers and very long times whose purpose is extreme adjacency effects and maximum apparent sharpness, said to be fraught with dangers, with high-speed films named as poor candidates, with the warning that even when it works one or more frames are often ruined by an air bubble or bromide drag, with the instruction to make two or more back-up shots of important scenes, a working dilution of about 2 parts A plus 2 parts B plus 400 to 500 parts water, 60 seconds of continuous vigorous agitation and then no further agitation, and times for most films in the 45 to 60 minute range. Fourth page - the statement that the contrast-index charts are based on rotary processing in BTZS type tubes with a five-minute pre-soak, and the table of optimum contrast index by printing processunblinkingeye.com/Articles/PCat/pcat.htmltier 2, specialist2026-09-06
- 07The Pyrocat-HD Developer, catalogue numbers 01-5080 and 01-5082 (dry) and 01-5081 and 01-5083 (liquid), to make 10 or 50 litres of working solution: technical informationPhotographers' Formulary, Inc.§ King's development recommendations - sheet film in trays with normal agitation at 10 seconds every minute; with minimal agitation at 10 seconds every three minutes and times about 50 per cent longer; and with semi-stand agitation at a special working solution of 1 part A to 1 part B to 200 or 400 parts water, agitation for one minute at the start of development followed by 30 seconds at the halfway point, times of 40 to 50 minutes for slow and medium-speed films and 50 to 60 for fast films, and a warning that dichroic fog may result from extended development of high-speed films; the rotary instruction to use a minimum of 75 mL of standard working solution per sheet of 4x5 film; and Ed Buffaloe's semi-stand account of two 35 mm rolls of Delta 3200 at 1:1:200 for 30 minutes with agitation for one minute at the start and 30 seconds in the middle, of which the first printed with extremely high acutance and the second, of a very high-contrast scene, was ruined because the bromides released by the intense development in the heavily exposed areas diffused out and caused uneven development in surrounding areasstores.photoformulary.com/content/01-5080.pdftier 1, primary2026-09-06
- 08Pyrocat-HD Film Developer: kit instructionsBostick & Sullivan, Inc.§ Introduction - the description of Pyrocat-HD as a semi-compensating high-definition developer and the claim of no streaking or mottling with reduced agitation, listed as an advantage over PMKbostick-sullivan.com/wp-content/uploads/2022/03/Pyro-HD-instructions.pdftier 1, primary2026-09-06
- 09BERGGER PMK DatasheetBERGGER, 2020§ Development - the instruction to agitate constantly for the first 15 seconds and then every 15 seconds, with the tank stationary between agitations, and the statement that this high stirring frequency prevents uneven development; Development errors - uneven development with areas of dissimilar density and uneven colouring listed against insufficient agitation, and denser image edges listed against inadequate shaking resulting in turbulence at the edges of the negative; and the development temperature range of 21 to 27 degrees C with the statement that emulsion damage follows above 27 to 28 degrees Cbergger.com/fr/index.phptier 1, primary2026-09-06
- 10Photographic Facts and FormulasE. J. Wall, F.C.S., F.R.P.S., 1924§ Development, Stand Development - the description of the method as a dilute developer in an upright grooved tank in which plates are left for a given time, the note that in the early days it was put forward as a cure for every error in exposure with solutions so dilute that 12 to 24 hours were required, the statement that its only advantages are fine grain and uniform results if the developer be frequently agitated, the report of Wratten and Wainwright's photometric measurements that a plate needing 3 minutes in rodinal at 1:20 needed 42 minutes at 1:200 in air-free distilled water, 46 in ordinary distilled water and 52 in tap water, the note that pyro-soda and glycin seem to be the only developers not affected by the water, and the statement that stand development is not economical because most commercial tanks require a large amount of solution which is as a rule so oxidised at the end of development as to be useless a second timearchive.org/details/photographicfact00walltier 1, primary2026-09-06
- 11Numerical investigation of sharpness in photographic layers containing DIR compounds, Optica Applicata volume XXIX number 3, pages 275 to 283Bogumil Rajkowski and Piotr Nowak, Institute of Physical and Theoretical Chemistry, Wroclaw University of Technology, 1999§ Introduction - the adjacency-effect mechanism at a knife-edge exposure, in which development-inhibiting by-products diffuse laterally from the high-exposure region into the low-exposure region while fresh developer diffuses the other way, so that development runs faster at the edge of the dense area and gives a local density maximum called the border effect and slower at the edge of the thin area and gives a local density minimum called the fringe effect; Experimental - the edge measurements made by scanning across knife edges with a microdensitometer, with the distance axis in micrometres; and the acutance definition, in which the starting and stopping points are set by the smallest visible density gradient, which Higgins and Jones found to be 0.005 density units per micrometredbc.wroc.pl/Content/40473/PDF/optappl_2903p275.pdftier 1, primary2026-09-06
- 12Kinetics, supplemental module 9: DiffusionLaura Dickson, University of California, Davis, for Chemistry LibreTexts§ Fick's first law stated as J equals minus D times the concentration gradient, with J the flux in mol per square metre per second; and the statement that ions at room temperature usually have a diffusion coefficient between 0.6 and 2 times 10^-9 square metres per secondchem.libretexts.org/Bookshelves/Physical_and_Theoretical_Chemistry_Textbook_Maps/Supplemental_Modules_(Physical_and_Theoretical_Chemistry)/Kinetics/09%3A_Diffusiontier 2, specialist2026-09-06
- 13Elementary Photographic ChemistryEastman Kodak Company, 1928§ Chapter III - the statement that the speed of development depends chiefly on the rate at which the developer diffuses into the filmarchive.org/details/elementaryphotog00east_0tier 1, primary2026-09-06
- 14Transmission Step WedgesStouffer Industries, doing business as Stouffer Graphic Arts§ Product table - the T2115, 21 steps at a nominal 0.15 density increment to a maximum density of 3.05; and the note that only the T2120CC and T1530CC are supplied calibratedstouffer.net/TransPage.htmtier 1, primary2026-09-06
- 15X-Rite 361T Desktop Transmission Densitometer, product brochure L11-010X-Rite, Incorporated§ Specification table - measuring areas of 1, 2 and 3 mm with 0.5 mm optional; repeatability plus or minus 0.01 D and linearity plus or minus 0.02 D from 0.0 to 5.0 D; cited only as what a commercial metal-cased instrument publishes about itselfxrite.com/-/media/xrite/files/literature/l11/l11-000_l11-099/l11-010_361t_product_brochure/l11-010_361t_en.pdftier 1, primary2026-09-06
- 16Uncertainty of Measurement Results (NIST Reference on Constants, Units, and Uncertainty)National Institute of Standards and Technology, Physical Measurement Laboratory, 2017§ Basic definitions - Type A evaluation of uncertainty as the statistical analysis of series of observations and Type B as evaluation by other means; and Combining uncertainty componentsphysics.nist.gov/cuu/Uncertainty/index.htmltier 1, primary2026-09-06
- 17ILFORD RAPID FIXER, technical informationHARMAN technology Limited (ILFORD Photo), 2010§ Film clearing time and the instruction to fix for twice the clearing time; fixing times for general purpose film at 1+4ilfordphoto.com/amfile/file/download/file/1833/product/711tier 1, primary2026-09-06
- 18ILFORD Chemical Sundries: ILFOSTOP, ILFOTOL and WASHAID, technical informationHARMAN technology Limited (ILFORD Photo), 2017§ The statement that a water rinse may be substituted for the stop bath but increases the risk of processing marks and stains; and ILFOTOL at 5 mL per litre in the final rinseilfordphoto.com/amfile/file/download/file/1865/product/669tier 1, primary2026-09-06
Formulas, hazard statements, historical dates and process descriptions on this page were checked against the sources above on the date shown. Safety data changes: obtain the current safety data sheet for the product you actually buy before you open it.