Uneven density across a strip
Lifted from the sensitometer enclosure and exposure stage, calibrating the sensitometer and the step-wedge exposure series, whose preparation section already contains the uniformity map this entry is the failure of.
What you see
Section titled “What you see”A step that is not one density. Read the same step near one edge of the strip and near the other and the two readings differ. On a long strip the difference usually runs one way — a gradient — and on a badly made enclosure it can be a bright patch instead.
The consequence is worse than the appearance. Every curve plotted from that strip inherits the gradient, and because the steps run along the strip, a gradient along the strip is indistinguishable from a change of slope. A sensitometer with a bright end quietly reports a contrast index that the film never had.
Likely causes
Section titled “Likely causes”- A non-uniform exposing field. The commonest cause on a home-built instrument, and the reason Part XIV’s enclosure page maps the field before anything else.
- Poor contact between the wedge and the emulsion. Light spreads under a lifted step and softens the boundary between steps as well as lifting the density.
- Uneven development. A strip lying against a tank wall, insufficient agitation, or a crowded tank — the same causes as mottled development and surge marks, acting on a piece of film small enough that the pattern is mistaken for a reading.
- The densitometer. An aperture placed differently on each step, on a strip whose steps are only half an inch wide, samples a different place each time.
What is happening: the chemistry and physics
Section titled “What is happening: the chemistry and physics”Nothing photographic. This is a geometry fault wearing a sensitometric costume, and that is precisely why it is dangerous: it produces a plausible curve.
A sensitometric exposure is only interpretable if the light arriving at the film is the same everywhere the wedge is not, because every density on the plot is attributed to the wedge step above it. Kodak’s workbook states both conditions — a uniform field over the area of the strip, and intimate contact between the modulator and the emulsion. Break either and the density you read is the sum of the step and something else, with no way to tell them apart from a single strip.
The development half is the same argument in chemistry. Development is a diffusion-limited process at the emulsion surface; where fresh developer reaches the film more readily, more silver is reduced. Part IX’s solvent and agitation experiments are where the course measured that, and a control strip is simply a very small negative subject to it.
Diagnostic questions
Section titled “Diagnostic questions”- Has the exposure field ever been mapped? Part XIV’s enclosure page calls this the step everybody skips. A field map is a one-off measurement that makes every later reading interpretable.
- Is the wedge in contact, and under what pressure? A glass plate, a spring back, or a vacuum are the three usual answers, and a wedge lying loose is none of them.
- Where in the tank was the strip, and what was the agitation? A strip taped to a wall develops differently from one hanging free.
- How large is the densitometer aperture relative to a step? A 21-step wedge is half an inch wide; an aperture near that size averages over a boundary.
- Does the same fault appear on a strip from a different film? If it does, it is not the film.
Corrective action
Section titled “Corrective action”Discard the strip. A gradient cannot be corrected out of a curve, because the correction and the signal have the same shape.
Fix the field before you fix anything else. An instrument that is not uniform makes every subsequent diagnosis unreliable, including the ones you would use to check the instrument.
If the fault is development, standardise the agitation as a written script — Part IX’s experimental design lesson is where the course argues for scripts — and process the strip in the same tank position every time.
Prevention
Section titled “Prevention”Map the field once and record the map. The calibration records worksheet exists for exactly this kind of measurement.
Give the strip its own tank position and its own place in the agitation script, so that whatever unevenness remains is at least constant.
Read the same place on every step, and record which place. A convention that is written down is a convention you can check.
Sources for this page
4 cited · checked 2026-09-05
- 01Basic Photographic Sensitometry Workbook, publication H-740Eastman Kodak Company§ Sensitometric exposure - the requirement that the exposing field be uniform over the area of the strip, and that the modulator be in intimate contact with the emulsionkodak.com/content/products-brochures/Film/Basic-Photographic-Sensitometry-Workbook.pdftier 1, primary2026-09-05
- 02Transmission Step WedgesStouffer Industries, doing business as Stouffer Graphic Arts§ Transmission step wedge specifications - the physical sizes of the wedges, the 21-step T2115 at half an inch by five inches, and the density increments of each typestouffer.net/TransPage.htmtier 1, primary2026-09-05
- 03Monitoring and Troubleshooting KODAK Black-and-White Film Processes, publication Z-133E, bound with How to Process and Print Black-and-White Film, publication AJ-3Eastman Kodak Company, 2005§ Z-133E - agitation as a named cause of uneven development in small tanks, and the instruction to check the measurement before changing the process when a point falls outside a control limit125px.com/docs/techpubs/kodak/z-133-2003_03b.pdftier 1, primary2026-09-05
- 04ISO 5-2:2009, Photography and graphic technology - Density measurements - Part 2: Geometric conditions for transmittance density, fifth edition, 2009-12-01ISO/TC 42 Photography and ISO/TC 130 Graphic technology, joint working group, 2009§ Density measurement, part 2 - geometric conditions for transmission density, and the influence of the measuring apertureiso.org/standard/52914.htmltier 1, primary2026-09-05
Formulas, hazard statements, historical dates and process descriptions on this page were checked against the sources above on the date shown. Safety data changes: obtain the current safety data sheet for the product you actually buy before you open it.