The calibration record set
Each sheet is the written half of a procedure that already exists. The steps are numbered here the way they are numbered there, so a sheet is filled in from top to bottom while the procedure is followed from step 1: CA-1 to CA-3 follow checking the balance and verifying the thermometer, and CA-4 and CA-5 follow calibrating a pH meter. CA-6, the graduate, is the third of the three commissioning checks, taught in commissioning your laboratory.
Print the instrument you are checking, not the set. A filled sheet goes into the calibration log at the front of the notebook, one instrument to a page, because ten entries on one page make a drifting thermometer visible and no single check can do that.
Printing, versions, and why nothing on a sheet explains anything: see printable worksheets.
Pure Silver · The calibration record set
CA-1
The balance — the session, the drift, and the ten readings
One balance, one check. Steps 1 to 3 of the procedure.
Sessionbefore
Step 2 — driftas you go
Step 3 — the ten readingsas you go
| Trial | Readingg | Anything you noticed |
|---|---|---|
| 1 | ||
| 2 | ||
| 3 | ||
| 4 | ||
| 5 | ||
| 6 | ||
| 7 | ||
| 8 | ||
| 9 | ||
| 10 |
Pure Silver · The calibration record set
CA-2
The balance — what the readings say, and the finding
The same check. Steps 3 to 12 of the procedure.
Step 3 — what the ten readings sayas you go
Step 4 — off-centre loadingas you go
| Position | Readingg | Difference from the centred meang |
|---|---|---|
| Centred, trial 1 | ||
| Centred, trial 2 | ||
| Near one edge, trial 1 | ||
| Near one edge, trial 2 |
The finding, and what is done about itafter
Pure Silver · The calibration record set
CA-3
The thermometer — the ice point
One thermometer, one verification. Steps 5 to 12.
Sessionbefore
Steps 5 to 7 — the ice pointas you go
| Trial | Reading°C | Reference0.0 °C | Deviation°C | Slurry still packed at the end |
|---|---|---|---|---|
| 1 | ||||
| 2 | ||||
| 3 |
Steps 8 and 9 — the second thermometer, and the boiling pointas you go
| Point | Liquid-in-glass°C | Probe°C | Difference°C |
|---|---|---|---|
| At the ice point | |||
| At room temperature |
The finding, and what is done about itafter
Pure Silver · The calibration record set
CA-4
The pH meter — the calibration
One meter, before one session. Steps 1 to 9.
Sessionbefore
The standardsbefore
| Standard | Lot | Date opened | Temperature°C |
|---|---|---|---|
| pH 7.00 | |||
| pH 4.01, for the calibration | |||
| pH 4.01, unopened until step 9 |
Steps 5 to 8 — the two pointsas you go
Step 9 — verification against a standard the calibration never sawas you go
Pure Silver · The calibration record set
CA-5
The pH meter — the session, and the closing drift
The samples that calibration was for. Steps 10 to 13.
Step 10 — the samples, in order of increasing pHas you go
| Sample | Time | Temperature°C | pH |
|---|---|---|---|
Steps 11 to 13 — closingafter
Pure Silver · The calibration record set
CA-6
The graduate
What one vessel delivers, by weighing the water it pours out.
Sessionbefore
The three trialsas you go
| Trial | Filled to the markmL | Mass of water deliveredg | Volume deliveredmL | Drain times |
|---|---|---|---|---|
| 1 | ||||
| 2 | ||||
| 3 |
The finding, and what is done about itafter
Pure Silver · The calibration record set
CA-7
The timer
One timer against a second clock. No limit is published for it.
Sessionbefore
The trialsas you go
| Trial | Interval set | Reference reading | Timer reading | Deviations | Deviationper cent of the interval |
|---|---|---|---|---|---|
| 1 | |||||
| 2 | |||||
| 3 |
The finding, and what is done about itafter
Pure Silver · The calibration record set
CA-8
Safelight fog test
One instrument or material, one dated measurement session.
Conditions and traceabilitybefore
Measurementsas you go
| Position / distance | Exposure duration | Control result | Exposed result | Difference |
|---|---|---|---|---|
Decisionafter
Pure Silver · The calibration record set
CA-9
Enlarger alignment and uniformity
One instrument or material, one dated measurement session.
Conditions and traceabilitybefore
Measurementsas you go
| Field position | Lens / aperture | Focus result | Relative illumination | Deviation |
|---|---|---|---|---|
Decisionafter
Pure Silver · The calibration record set
CA-10
Sensitometer repeatability
One instrument or material, one dated measurement session.
Conditions and traceabilitybefore
Measurementsas you go
| Run | Exposure setting | Reference response | Measured response | Deviation |
|---|---|---|---|---|
Decisionafter
Pure Silver · The calibration record set
CA-11
Densitometer calibration
One instrument or material, one dated measurement session.
Conditions and traceabilitybefore
Measurementsas you go
| Reference density | Dark reading | Reference reading | Sample reading | Computed density |
|---|---|---|---|---|
Decisionafter
Where the fields come from
Every field on these sheets is one a written page of the course asks you to record. Those pages are linked here, and the build fails if one of them stops existing.
- Procedure — balance and thermometer checkThe procedure sheets CA-1 to CA-3 follow, step for step, and the log line it ends with.
- Procedure — ph meter calibrationThe two-point calibration, the verification against a standard the calibration never saw, the closing drift check, and the log line.
- Part 2 — commissioning your laboratoryFive tables of instrument checks — the balance ten times over, the thermometer at its ice point, and what the graduate actually delivers.
- Part 2 — measurement and uncertaintyThe three checks you can actually perform, and why repeatability and trueness are different findings.
- Part 3 — measuring phRuns the meter's calibration, and keeps the calibration log at the front of the notebook.
- Part 17 — calibrating the timerArm 2, the timer against a second clock over an hour, and the timer certificate the arm's result is filed on.
- Part 16 — the safelight fog testRecords exposed and control material under the measured safelight arrangement.
- Part 16 — commissioning the enlargerRecords alignment, focus and illumination across the image field.
- Part 14 — calibrating the sensitometerRecords repeated exposure checks and the declared repeatability criterion.
- Part 15 — calibrating the densitometerRecords reference readings, calibration, geometry and residuals.
What these sheets do not carry
- Acceptance limits and review intervals remain declared by the student against the cited method; the course does not invent a domestic instrument standard.